Product Description
Evaluates oxidation stability by subjecting the sample to a temperature of 95°C in the presence of oxygen or dry air. For inhibited mineral oils, the sample is reacted with oxygen in the presence of water and an iron-copper catalyst.
~ Conforms to ASTM D943, D2274, D4310, and related specifications
~ 8 and 12 position benchtop liquid baths available
~ 30 and 60 position floor model liquid baths available
~ Custom configurations for D2983 and AOCS CD12-57 tests available
Constant temperature bath with solid state temperature control, calibrated flowmeters and condenser water manifold for oxidation stability tests on fuels and lubricants. Individual flowmeters and control valves for each oxidation cell deliver air flow at the rate of 3L/h. Condenser water manifold has individual control valves for each cell. Temperature controller provides quick temperature stabilization without overshoot, and the bath is protected by a redundant overtemperature control circuit that interrupts power should bath temperature exceed a programmed cut-off point. Display provides actual setpoint temperature values in °C/°F format. Communications software (RS232, etc.), ramp-to-set and other enhanced features are available as extra cost options. Double-wall insulated baths are equipped with copper immersion heaters and a 1/20 hp circulation stirrer. Stainless steel bath interior has a builtin support rack and overflow/drain to immerse the test cells at the required depth. Order oxidation cell glassware and accessories separately.
SPECIFICATIONS:
Capacity: 8 Oxidation Cells
Temperature:
~ Range: Ambient to 250°F (121°C)
~ Control Stability: ±0.09°F (±0.05°C)
Bath Medium: 9.9 gal (37.5 L) white technical oil
~ Low-metalloid steel wire and electrolytic copper wire wound in a double spiral conforming to ASTM specifications.
~ Packed in a sealed glass tube with a nitrogen atmosphere. Ready for use.
~ K24000 Wire Coiling
~ Mandrel Mounts on bench for winding steel and copper wire into catalyst coils meeting ASTM specifications.
~ K12250 Steel Wire
~ Low metalloid steel wire, 0.0625 in. (1.59mm) diameter, for catalyst coils.
~ Supplied in 1000 ft (304.8m) lengths.
~ K12260 Copper Wire
~ Electrolytic copper wire, 0.064 in. (1.63mm) diameter, for catalyst coils.
~ Supplied in 1000 ft (304.8m) lengths.
~ 380-100-001 Silicone Carbide Paper
~ Used to polish steel and copper wire prior to winding into catalyst coils. 100 grit.
~ 250-002-001 Oxidation Cell Thermometer - Range: 80 to 100°C. For ASTM D943 and D4310.
~ 250-000-40C ASTM 40C Thermometer - Range: 72 to 126°C. For constant temperature baths.
_
The following link provides additional information from the manufacturer's brochure. Click to open the PDF in a new window or right-click to download it:
Koehler K64200 / K64290 8-place Oxidation Stability Tester or copy the following into your browser address window: labequip.com/stock/pictures/n2227.pdf
Made in USA.
ACCESSORIES:
OPTIONAL ACCESSORIES AVAIABLE AT ADDITIONAL COST:
- K12281 Oxidation Cell Assembly for ASTM D943 and D4310 - Includes oxidation cell, condenser, oxygen delivery tube, thermometer bracket, oil level indicator strip, syringe sampling tube, sampling tube holder, spacer, PTFE stopper and O-rings
- K122-0-18 Oxygen Delivery Tube
- K122-0-19 Oxidation Test Tube
- K122-0-20 Condenser
- K122-0-21 Thermometer Bracket
- K122-0-22 Oil Level Indicator Strip
- K122-0-23 Syringe Sampling Tube Holder
- K122-0-27 PTFE Stopper
- K122-0-28 Syringe Sampling Spacer
- K122-0-30 Syringe Sampling Tube
- AS568-009-V14 O-rings
- K12210 Catalyst Coil - Low-metalloid steel wire and electrolytic copper wire wound in a double spiral conforming to ASTM specifications. Packed in a sealed glass tube with a nitrogen atmosphere. Ready for use.
- K24000 Wire Coiling - Mandrel Mounts on bench for winding steel and copper wire into catalyst coils meeting ASTM specifications.
- K12250 Steel Wire - Low metalloid steel wire, 0.0625" (1.59mm) diameter, for catalyst coils. Supplied in 1000 ft (304.8m) lengths.
- K12260 Copper Wire - Electrolytic copper wire, 0.064" (1.63mm) diameter, for catalyst coils. Supplied in 1000 ft (304.8m) lengths.
- 380-100-001 Silicone Carbide Paper - Used to polish steel and copper wire prior to winding into catalyst coils. 100 grit.
- 250-002-001 Oxidation Cell Thermometer - Range: 80 to 100°C. For ASTM D943 and D4310.
- 250-000-40C ASTM 40C Thermometer - Range: 72 to 126°C. For constant temperature baths.